Pacemakers and ICD

Indications for emergency pacing

  • Symptomatic bradycardia
  • Symptomatic high-grade AV block (Mobitz II or complete)
  • Severe sick sinus syndrome with asystole >3s and syncope
  • Overdrive pacing for torsades de pointes
  • Overdrive pacing for recurrent monomorphic VT
    • Risk of inducing VF and limited by machine pacing limit of 180
  • Overdrive pacing of unstable SVT
    • Only once pharmacological and electrical cardioversion have failed

Emergency pacing

  • Transcutaneous pacing for severe hypotension with bradycardia
    • 70% survive with good neurological outcome vs. 15% with non-pacing modalities
    • May be able to pace RV infarct when transvenous pacing fails (as can directly pace LV)

Technique for transcutaneous pacing

  • External pads and electrodes for monitoring required
  • If bradyasystolic peri-arrest, ramp up current to 100mA and titrate down once stabilised
  • If less severe compromise, can titrate up from 10mA to usually 50-100mA and then maintain at 1.25x the threshold
  • Set rate to 80bpm and increase by 10bpm until perfusing (up to 100/min)
  • May be fixed or demand (synchronous)
    • Fixed (asynchronous) pacing carries risk of R on T, however, there is little outcome data to support a preference for either
  • Need electrical and mechanical capture
    • US can be helpful for ensuring mechanical capture if poorly perfused
  • Reasons for failed capture may include
    • Inadequate current
    • Faulty electrical contact
    • Electrode placement
    • Patient size
    • Underlying pathology
    • PTX, severe ischaemia, pericardial effusion or metabolic derangement

Resuscitation in patients with permanent pacemakers

  • If countershock required, place pads at least 8cm from the pulse generator
  • After countershock, interrogate the pacemaker to ensure still functioning as can suffer from:
    • Pacemaker inhibition due to reversion to noise mode
    • Deletion (reprogramming)
    • Circuit damage
    • Myocardial damage near lead tip caused by current transmission via the electrode to the myocardial interface
  • Defibrillation can lead to global myocardial ischaemia which in turn also increases the pacing threshold thus causing capture failure
    • If this occurs, try transcutaneous pacing at a higher current

Nomenclature

Letter positionIIIIIIIVV
CategoryChamber(s) pacedChamber(s) sensedResponse to sensingProgrammability, rateAntiarrhythmic functions

O, none A, atria V, ventricle D, dual (A+V) S, single chamberO, none A, atria V, ventricle D, dual (A+V) S, single chamber
O, none T, triggered I, inhibited D, dual (D + I)O, none P, simple program M, multiprogram C, communicating (telemetry) R, rate modulationO, none P, pacing S, shock D, dual (P+S)

Most commonly VVI and DDD

The fifth letter is rarely used

Third position D indicates both triggered and inhibition responses can occur

i.e. If atrial activity sensed, atrial pacing is inhibited but triggers ventricular pacing

Programmability

  • Activity sensors: Vibration detectors (accelerometers or piezoelectric crystal)
    • May respond to certain activities e.g. using a drill
  • Minute ventilation sensors
    • Measures impedence between pacemaker unit and electrode

Magnet inhibition

  • Placing magnet over permanent pacemaker causes sensing to be inhibited and results in asynchronous mode (AOO, VOO, DOO)
    • Usually at rate of 100
    • Risk of R on T

Pacing modes for bradycardia

  • Single-chamber pacing
    • AOO and VOO
      • Asynchronous and virtually obsolete except in a few emergency situations
    • AAI
      • Atrial demand pacing
      • Indicated in sinus bradycardia providing AV conduction is intact
      • Only paces if does not sense atrial activity
    • VVI
      • Ventricular demand pacing
      • Most commonly used mode in life-threatening bradycardias
      • Spontaneous ventricular activity is sensed and low risk of R on T
      • AV synchrony is lost

Pacing modes for bradycardia

  • Dual chamber pacing
    • DVI (AV sequential pacing)
      • Will pace atria, wait and then pace ventricle if no ventricular activity sensed
      • To maintain AV synchrony in the absence of atrial sensing, pacing rate must be higher than intrinsic spontaneous atrial rate
      • Indicated if impaired AV conduction and atrial bradycardia
      • Not useful if atrial tachyarrhythmias exist as cannot pace faster than them
    • VDD (Atrial synchronous ventricular inhibited)
      • Paces only the ventricle but senses both
      • Sensed P-wave triggers ventricular pacing

Pacing modes for bradycardia

  • Dual chamber pacing
    • DDD (Dual pacing and sensing)
      • Atrial impulse triggers ventricular pacing unless senses autonomous ventricular activity
      • Upper rate limiters prevent ventricular pacing of atrial tachyarrhythmias
      • If atrial bradycardia with intact AV conduction – Atrial pacing
      • If sinus rhythm with AV block – synchronised ventricular pacing of intrinsic P waves
      • If sinus bradycardia with AV block – Sequential atrial then ventricular pacing
      • Normal sinus rhythm and AV conduction – Inhibition of both atrial and ventricular pacing
      • Can suffer re-entry loops where PVC is transmitted retrogradely to atria, where it is sensed and results in ventricular pacing with endless loop

Pacing modes for bradycardia

  • Dual chamber pacing
    • DDI (AV sequential, non-P-wave synchronous)
      • Sensing of both but sensed atrial events do not trigger ventricular pacing
      • Prevents endless loop phenomenon or tracking of SVTs
      • Useful for SA node dysfunction with episodic atrial tachyarrhythmias
      • Will simply pace ventricle at backup rate in the setting of atrial tachyarrhythmia

Complete AV block in acute MI

FeatureInferiorAnterior
OnsetSlowSudden
TypeMobitz IMobitz 2
Ventricular rate>45<45
Escape pacemakerStableunstable
Response to atropineYesNo
Haemodynamic effectsNoYes
Permanent pacingNoYes (if high-degree AV block)
PrognosisGoodTerrible

Indications for permanent pacing

  • Class I
    • Chronic symptomatic 2nd or 3rd degree AV block
    • SA node dysfunction with documented sinus bradycardia
    • Recurrent syncope associated with carotid sinus hypersensitivity
  • Class II-IIa
    • Asymptomatic complete AV block with average ventricular rate >40 in awake patient
  • Class IIb (weak supportive evidence)
    • 1st degree block with depressed LV function and symptoms of LV failure
  • Class III (not indicated)
    • Asymptomatic 1st degree block or reversible AV block due to drugs

Indications for permanent pacing

  • Other
    • HOCM (Class IIb)
      • Usually for symptomatic patients with high gradient in LVOT
      • DDD pacing with short AV interval caused RV apical activation with altered septal activation to reduce LVOT gradient and systolic anterior motion of mitral valve
    • Heart failure
      • Class IIa for NYHA III/IV patients with dilated or ischaemic cardiomyopathy, QRS >130ms
      • Cardiac resynchronisation therapy (CRT) is indicated in these patients via biventricular pacing to improve survival and symptoms

Fixed-rate vs. demand

  • Fixed-rate pulse generators produce an electrical signal regardless of patients own intrinsic electrical rhythm
    • Can result in serious arrhythmias if discharges during vulnerable period (T wave)
  • Demand pacing
    • Typically discharges if no sensed electrical activity after certain time period 
    • May be inhibited by intrinsic sensed electrical activity or triggered to discharge during absolute refractory period

Pacing in tachyarrhythmias

  • May be useful for:
    • SVT: AVNRT, AVRT (rarely required)
    • Atrial flutter (rapid atrial overdrive pacing)
    • Unifocal atrial tachycardia (rapid atrial overdrive pacing helpful if re-entry pathway driven but less so if autonomous focus firing at rapid rate)
    • Ventricular tachycardia
      • Should not be used if very rapid ventricular rates >300 or significant haemodynamic instability exists
  • No value in sinus tachycardia, AF or VF

Pacing in tachyarrhythmias

  • Torsades
    • Pace atria or ventricle at 110-120/min
  • SVT
    • Atrial pacing at 60-80 and slowly increase to 10-20% faster than spontaneous atrial rate
    • Atria then paced for around 30 seconds then switched off with ensuing sinus rhythm
    • If fails, try different atrial pacing site and faster rate

Pacing in tachyarrhythmias

  • VT
    • Ventricular burst pacing
      • Pace ventricle at 120% of spontaneous VT rate for 5-10 beats then stop
      • Can precipitate faster VT and VF
    • Underdrive ventricular pacing at rate <50% of VT rate is sometimes successful
    • Overdrive atrial pacing may be useful if 1:1 AV conduction and ventricular rate relatively slow (120-180/min)

Features

  • Lithium batteries have lifespan of 8-12 years
  • Most units preset for rates near 70, with pacing interval of 0.84 seconds
  • Demand pacemakers have built-in refractory period of 0.2 to 0.4 seconds during which it will not sense, to prevent it being inhibited by its own stimulus
  • Magnets held over most units will convert them from demand to fixed-rate mode
    • Can quickly ascertain paced rate, however, should only be performed for short periods due to risk of arrhythmia in fixed-rate mode (due to stimulus in vulnerable period)
  • More sophisticated units can be interrogated 

Pacemaker malfunctions

  • 1) Problems with the pocket
  • 2) Problems with the leads
  • 3) Failure to pace
  • 4) Failure to sense – Leads to fixed rate pacing and risk of arrhythmia
  • 5) Malfunction causing overpacing or runaway pacing

Pacemaker evaluation

  • Evaluation
    • Examination of pocket
    • CXR to confirm lead placement and device itself
    • Electrolytes
    • Cardiac enzymes
    • ECG
    • Interrogation
    • Magnet fixed-rate testing

Pacemaker syndrome

  • Seen in 20% of patients in early phase with symptoms of syncope, near-syncope, orthostatic dizziness, exercise intolerance, dizziness, uncomfortable pulsations over neck/abdomen and RUQ pain
  • AV synchrony and presence of ventriculoatrial conduction are most common in VVI but can be seen with DDI mode
  • In VVI
    • If the sinus node is intact, an atrial impulse can occur when the tricuspid and mitral valves are closed with increased jugular and pulmonary venous pressure leading to symptoms of CCF
    • Atrial distension can lead to reflex vasodepressor effects mediated by the CNS
    • If contribution of atrial contraction to late diastolic ventricular filling is important, then orthostatic hypotension can occur
  • In DDI with AV block
    • If sinus node discharge rate exceeds the programmed rate of the pacemaker, can get atrial contraction against closed valves with subsequent pacemaker syndrome
  • Typically symptoms are mild and patients adapt to them over time
  • Unfortunately, symptoms are severe in 1/3 and may warrant upgrade from VVI to dual-chamber pacing or lowering the rate of VVI pacing 
  • If symptoms occur in DDI, then optimising the timing of atrial and ventricular pacing is often required

Pacemaker infection

  • Seen in <1% soon after placement
  • Mostly S. aureus or S. epidermidis early on
  • Presents as local inflammation or abscess at site
  • Skin adherence to the device with discolouration is highly suggestive
  • If only superficial infection suspected, antibiotics and analgesics with early review is indicated
  • If erosion of skin occurs, requires surgical replacement
  • Cardiac device-related IE can occur presenting with sepsis and positive BC without local inflammation at pocket

Thrombophlebitis and venous obstruction

  • Extensive venous collaterals exist making this quite rare (0.3-3% of patients)
  • Site of insertion makes no difference
  • Symptoms include oedema, pain or venous engorgement ipsilaterally
  • Treatment is IV heparin then oral NOAC/warfarin

Pneumothorax

  • Seen in 1% of patients after insertion
  • More common with subclavian insertion technique

Undersensing

  • Failure to sense intrinsic cardiac activity with subsequent asynchronous, fixed-rate pacing
  • Causes include increased threshold (exit block), poor lead contact, new BBB, inferior MI or programming issues
  • Pacing spikes within QRS complexes is highly suggestive of this
  • Lead dislodgement usually occurs within 2 days of insertion

Oversensing

  • Pacemaker senses electrical activity not from atria or ventricles; it is thus inhibited, and generation of pacemaker impulse suppressed leading to bradycardia
  • Unipolar electrodes are more prone to environmental sensing than bipolar leads
  • Suxamethonium-induced depolarisation fasciculations
  • Crosstalk (atrial output sensed by ventricular lead)
  • Why might tachyarrhythmia arise?
    • Intrinsic depolarisation occurring during pacemaker refractory period, this not being sensed, and pacemaker firing soon after in the vulnerable period to initiate a re-entrant tachycardia
    • Maintenance of the tachyarrhythmia does then not involve the pacemaker at all
    • Emergency therapy can involve re-programming, or more commonly, magnet conversion to asynchronous, fixed-rate mode

Output failure

  • Paced stimulus not generated when expected
  • Results in decreased or absent pacemaker activity on ECG
  • Causes include oversensing, wire fracture, lead displacement
  • Steps:
    • Make sure pacemaker box is on and connected
    • Increase pacemaker current (up to 200mA for transcutaneous or 20mA for transvenous)
    • Asynchronous DOO/VOO mode selected to avoid oversensing
    • Convert to transcutaneous pacing  while new pacing system inserted

Failure to capture

  • Paced stimulation does not cause myocardial depolarisation
  • Electrode displacement, wire fracture, electrolyte disturbance, MI at lead tip or exit block
  • Seen with amiodarone, flecainide, hyperkalaemia, acidosis, alkalosis, cardiac perforation and improper settings also
  • If patients native heart rate is higher than the paced rate, no pacemaker activity is expected and output failure/capture failure cannot be recognised on ECG
  • Suxamethonium causes depolarisation fasciculations that can lead to oversensing and subsequent failure to capture

Failure to capture

  • Rx
    • Place in left lateral position (maximises electrode contact with endocardium)
    • Maximise pacemaker output
    • Use asynchronous mode
    • Transcutaneous if failing transvenous (as can directly pace LV)
    • Isoprenaline/adrenaline infusion to increase intrinsic HR

Pacemaker-associated dysrhythmias

  • Pacemaker-mediated tachycardia (PMT)
    • Re-entry tachycardia with antegrade pathway via pacemaker and retrograde via AV node
    • Caused by retrograde P waves being sensed as native atrial activity by dual chamber pacemaker
    • Get paced tachycardia at rate determined by pacemaker
    • Can be terminated by AV block e.g. adenosine OR magnet application
  • Sensor-induced tachycardia
    • Modern sensors respond to exercise, tachypnoea.etc.
    • Sensors may misfire in the presence of vibrations, loud noise, fever, limb movement or electrocautery with subsequent inappropriately fast rate
    • Ventricular rate cannot exceed programmed max of 160-180
    • Terminated by magnet application

Pacemaker-associated dysrhythmias

  • Runaway pacemaker
    • Low battery voltage in older units results in pacing spikes at 2000bpm with risk of subsequent VF
    • Paradoxically, may be failure to capture due to low amplitude pacing spikes and subsequent bradycardia
    • Application of magnet may be lifesaving
  • Lead displacement arrhythmia
    • If dislodged, lead may float in RV causing tickling of myocardium causing ventricular ectopics, possible VT and failure to capture
    • CXR helps confirm diagnosis
    • If changes from typical LBBB pattern (indicating RV placement) to RBBB pattern, suggests erosion through interventricular septum

Pacemaker-associated dysrhythmias

  • Twiddler’s syndrome
    • Pulse generator rotation in pocket with dislodgement of leads and subsequent diaphragmatic or brachial plexus pacing
  • RV perforation
    • RBBB instead of LBBB
    • Pacing of diaphragm
    • Haemopericardium rarely

ICD

  • Reduce mortality from 30-45% to <2% per year in those at risk of sudden cardiac death
  • Mostly follow a tiered approach to ventricular arrhythmias
    • Antitachycardia pacing
    • Low-energy cardioversion
    • Defibrillation
  • Projected lifespan of 6-9 years
  • Most common cause of death is CCF and this should be managed as usual

ED evaluation of ICD

  • Causes of inappropriate shock delivery
    • False sensing
    • SVT with RVR
    • Muscular activity (shivering, diaphragm contraction)
    • Extraneous source (tapping of chest wall, vibrations)
    • Sensing T waves as QRS (double counting)
    • Sensing lead fracture or migration
    • Unsustained tachyarrhythmia
    • ICD-pacemaker interactions
    • Component failure

ED evaluation

  • Ask about symptoms around event, number of shocks, activity at the time and any recent anti-arrhythmic drug changes
  • Look for signs of trauma
  • 12-lead
    • Any shock-related ST changes should resolve within 15 minutes and if not suggests new ischaemia
  • CXR
    • Electrode migration, displacement or fracture
  • Anti-arrhythmic drug levels and serum electrolytes
  • If patient is receiving repeated inappropriate shocks, temporarily deactive with a magnet over the device (can simply remove magnet if want to shock again)
    • All ICD’s should then be evaluated by a cardiologist if been magnetised
  • If in cardiac arrest:
    • Follow normal protocols
    • Place pads at least 8cm from generator

Disposition

  • Need to discuss with treating cardiologist
  • Admission generally warranted if:
    • Cardiovascular instability
    • 2 or more shocks in a 1-week period
    • Correctable causes of dysrhythmia
    • Any sign of infection or mechanical disruption of the system

Last Updated on October 28, 2020 by Andrew Crofton